Chapter 14 specifications, 1 specifications of dr232/dr242 and ds400/ds600, Specifications of dr232/dr242 and ds400/ds600 -1 – Yokogawa DR240 User Manual
Page 246
14-1
IM DR232-01E
Specifications
14
14.1 Specifications of DR232/DR242 and DS400/
DS600
Style Numbers DR232/DR242: S8
DS400/DS600: S8
Connecting Subunits to the DR232/DR242
Types of Subunits
Subunit DS400: four modules
Subunit DS600: six modules
Number of Subunits which can be connected
Up to six units can be connected (either DS400 or DS600)
Connection method
Using dedicated cables
Available Modules
Main Unit
Communication interface modules : Selectable from GP-IB and RS-232-C
DI/DO modules
: Alarm 2-ch output (transfer contact), fail output module,
chart-end output and input signal for remote control (12
points)
Alarm modules
: 4-ch output(transfer contact) or 10-ch output (make contact)
• Number of modules which can be connected: 1
Communication module + DI/DO module + alarm module: not more than one
Subunit (DS400/DS600)
Input modules
: universal (DC voltage, thermocouple, RTD, contact), DCV/TC/DI
Alarm modules
: 4-ch output (transfer contact) or 10-ch output (make contact)
DI/DO modules : Alarm 2-ch output (transfer contact), fail output module, chart-end output and
input signal for remote control
• Number of modules which can be connected to one DS400: 4
Input module + alarm module + DI/DO module: max. four
• Number of modules which can be connected to one DS600: 6
Input module + alarm module + DI/DO module: max. six
* Only one DI/DO module can be connected to one system.
Input
Measurement range
Refer to 14.2 on page 14-7.
Measurement interval
Selectable from 0.5, 1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30, 60s; max. 500ms/300ch
(with sub-unit connected, filter: OFF, A/D integration period of 20ms or 16.7ms)
A/D integration time
Selectable from 20ms (50Hz), 16.7ms (60Hz), 100ms (10Hz) selectable, or automatically
switched between 50/60Hz. (Except, auto switch does not function on the subunit (DS400/
DS600) of a DC power supply model. Selecting “AUTO” will set the A/D integration time to 20
ms (50 Hz).)
Recording
Recording method
Raster scan method, 10-color wire dot recording
Number of recording points
Measurement results:
Max.300 points
Recording paper
Ordinary recording chart, Z-fold chart (total width 342.5mm, length 30m)
Effective recording width: 250mm (for dot recording)
Recording colors
Analog trend mode
Dot recording: purple, red, green, blue, brown, black, navy blue, yellowish green, redish purple,
orange (can be specified for each channel)
Digital printout: black
Alarm printout: red (however, alarm release mark: blue