Init balance, Indication of the status of initial balancing, Alarm ack init_balance – Yokogawa DR240 User Manual
Page 115: Balance mode = exec, Balance mode = init
5-12
IM DR232-01E
5.5 Performing Initial Balancing/Initialization on a
Strain Input Channel
D I S P
M O D E
P R I N T
F U N C
I N S
D E L
R A N G E A L A R M
S E T
C H A R T
E N T E R
F E E D
M. F U N C 1
M. F U N C 2
RECORD
ALARM
CHART
KEYLOCK
R E C O R D
INIT BALANCE
>Select Function item
E S C
• Press the FUNC key to enter the FUNC menu.
• Using
and
, select and/or enter a value for each of the shaded fields shown below.
• To exit any of the following menu items during the procedure, press the MODE (ESC) key.
This returns to the first item of the menu. Note, however, that your new settings and selections
are canceled.
• If you have made changes to any item executable with the FUNC key, information on initial
balancing may not appear on the display.
Performing Initial Balancing
ALARM ACK
INIT_BALANCE
(Operating status)
ENTER
(Main menu item)
BALANCE CH=001-010
>Select Channel No.
BALANCE MODE = EXEC
ABORT EXEC INIT
Balancing... *
> CHANNEL = 001-60
>OK
001
>OK
031
ENTER
ENTER
ENTER
Initializing the Strain Input Channels
ALARM ACK
INIT_BALANCE
(Operating status)
ENTER
(Main menu item)
BALANCE CH=001-010
>Select Channel No.
BALANCE MODE = INIT
ABORT EXEC INIT
ENTER
ENTER
Indication of the Status of Initial Balancing
If you have performed initial balancing, the DR recorder shows whether the initial balancing is
successful or not in the following format. Using the $ and $ keys, you can change the unit
number for which you want the status shown.
> CHANNEL = 001-60
>OK
001
>OK
031
Module-by-module status indication
: successful end of initial balancing
: unsuccessful end of initial balancing
Starting channel
Note
• This function is disabled during making a report.
• The following operations are stopped when this function is performed.
• Recording
• Saving data to the memory and reading data from it.
• Computing