Yokogawa DR240 User Manual
Page 266
14-21
IM DR232-01E
Specifications
14
Accuracy of Bridge Resistors
±0.01%, ±5 ppm/˚C (excludes contact resistance of the jumper setup switch)
Contact Resistance of the Jumper Setup Switch
100 m
Ω maximum
Input Resistance
10 M
Ω minimum
Filter
Third-order sinc filter, moving-average
Withstanding Voltage
Across channels: 50 V DC (except for DU500-14)
Across output terminals and ground: 1,500 V AC, 1 min (excluding DU500-14)
Normal Operating Temperature/Humidity Range
0˚ to 50˚C
20 to 80% RH for an ambient temperature range of 0˚ to 40˚C; 10 to 50% RH for 40˚ to 50˚C
(non-condensing)
Power Consumption
Included in the value of a system's installed main unit or sub-unit
Dimensions
DU500-12/DU500-13: Approximately 114.3 (W)
× 137 (H) × 88 (D) (mm)
DU500-14: Approximately 114.3 (W)
× 137 (H) × 77 (D) (mm)
Weight
DV 500 -12, -13
0.6kg
DV 500 -14
0.8kg
Installation Category (Overvoltage Category)
CAT II (IEC1010)
Accessory
Bridge box: 319300 (120
Ω)
14.5 Specifications of Strain Input Module