Yokogawa DR240 User Manual
Page 253
14-8
IM DR232-01E
Communication Function
Communication interface is possible by GP-IB, RS-232-C, RS422-A/RS-485, or ethernet.
Power Supply
Rated supply voltage
AC power supply : 100 to 240 VAC (freely selected)
DC power supply : 12 to 28 VDC (only for the subunit (DS400/DS600) of a DC power supply
model)
Operating supply voltage
AC power supply : 90 to 250 VAC
DC power supply : 10 to 32 VDC (only for the subunit (DS400/DS600) of a DC power supply
model)
Rated supply frequency
50/60 Hz
Power consumption
Main unit: max. approx. 130 VA
Sub-unit DS400: max. approx. 55 VA (AC power supply), max. approx. 25 VA (DC power
supply)
Sub-unit DS600: max. approx. 70 VA (AC power supply), max. approx. 25 VA (DC power
supply)
Fuse Ratings
• Main unit
Maximum rated voltage/current: 250 V/2.5 A, Type: Time-lag, Standard: IEC/VDE
• Subunit (The fuse may not be replaced by the customer.)
AC power supply
Maximum rated voltage/current: 250 V/2 A, Type: Time-lag, Standard: IEC/VDE
DC power supply
Maximum rated voltage/current: 250 V/6.3 A, Type: Time-lag, Standard: UL/CSA
Normal Operating Conditions
Supply voltage
AC power supply : 90 to 250 VAC
DC power supply : 10 to 32 VDC (only for the subunit (DS400/DS600) of a DC power supply
model)
Supply frequency
50Hz
±2%, 60Hz ±2%
Ambient temperature
Main unit : 0 to 50
°C
DS400/DS600 Subunit :
when floor-mounted : 0 to 50
°C , when panel-mounted : –10 to 60°C
However, if the subunit DS400/DS600 is a DC power supply model, it can be used in the range
from 0
°C to 50°C regardless of the mounting method.
Ambient humidity
20 to 80%RH for –10 to 40
°C, 10 to 50%RH for 40 to 50°C, 5 to 30%RH for 50 to 60°C (no condensation)
Vibration
10 to 60 Hz, 0.2m/s
2
Shock
Not allowed
Magnetic field
400 A/m max. (50/60Hz)
Position
Unit should be positioned left-right horizontally
Installation location
Room
Installation height
Altitude up to 2,000 m
Installation category based on IEC 1010-1
II*
1
Pollution degree based on IEC 1010-1
2*
2
Warm-up time
At least 30 minutes after power switch-on.
*1 Installation category is the specification of the inpuls withstanding voltage which is also called as
overvoltage category.
*2 Polution degree is the level of foreign body adhesion such as the solid, liquid, and gas which decrease
the wuthstanding voltage, 2 means general indoor atmosphere.
14.1 Specifications of DR232/DR242 and DS400/DS600