Trigger types – Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 88
LabMaster 10 Zi Oscilloscopes
80
922561-00 Rev A
Trigger Types
A set of standard trigger types is available for selection. These triggers are
basic waveform features or conditions such as a positive or negative slope
and are available on all oscilloscopes, along with tools like Software
Assisted Trigger and Trigger Scan.
There are more sophisticated triggers for setting complex conditions. These
include Smart Triggers, Serial Trigger, Measurement Triggers, and
MultiStage Triggers.
Edge
A simple trigger, Edge trigger is activated by basic waveform
features or conditions such as positive or negative slope, and
holdoff.
NOTE
: Up to 15 GHz bandwidth is available while Edge Triggering (provided no
holdoff is selected).
Width
Width trigger allows you to define a positive- or negative-going
pulse width bounded by a voltage level, above or below an
occurring trigger. Or you can specify a pulse width and voltage
range, within or outside an occurring trigger.
Pattern (Logic)
Logic trigger enables triggering on a logical combination
(pattern) of five inputs: CH1, CH2, CH3, CH4, and AUX
(sometimes referred to as EXT on legacy oscilloscope models).
You have a choice of four Boolean operators (AND, NAND, OR, NOR), and
you can stipulate the high or low voltage logic level for each input
independently.