Power

AC Power Source

100 to 240 VAC (±10%) at 50/60 Hz (± 10%)

Manual voltage selection is not required because the instrument
automatically adapts to line voltage.

Power Consumption

MCM-Z

I

M

ASTER

C

ONTROL

M

ODULE

Max. consumption (all accessories installed): ≤ 450 watts (450 VA)
Standby consumption: 5W

A

CQUISITION

M

ODULES

20-36 GHz Models

Max. consumption (all accessories installed): ≤ 1225 watts (1225 VA)
Standby consumption: 12W

50-65 GHz DBI Models

Max. consumption (all accessories installed): ≤ 1275 watts (1275 VA)
Standby consumption: 12W

Power and Ground Connections

The MCM-Zi Master Control Module is provided with a 10A/250V rated
grounded cord set containing a molded three-terminal polarized plug with a
standard IEC-60320 (Type C13) connector. The connector mates to a
compatible power inlet on the Master Control Module for making line
voltage and safety ground connections.

The 10-xxZi Acquisition Module is provided with a 16A/250V or 15A/125V
rated grounded cord set containing a molded three-terminal polarized plug
and a specific IEC-60320 (Type C19) connector. The connector mates to a
compatible power inlet on the Acquisition Module for making line voltage
and safety ground connections.

The AC inlet ground is connected directly to the frame of the instrument.
For adequate protection again electric shock, connect to a mating outlet
with a safety ground contact.