Getting Started Manual

922561-00 Rev A

63

4.

Touch the Enable Timeout checkbox.

5.

Touch the Timeout data entry control and provide a timeout value.

NOTE

: Use the sequence mode timeout to automatically interrupt the

sequence acquisition if the timeout value is exceeded without a valid trigger.
The timeout period accounts for instances when a

Num Segments

miscount occurs for some reason and the oscilloscope waits indefinitely for
an unforthcoming segment. During that time, no oscilloscope functions are
accessible. By means of a timeout value, however, the acquisition will be
completed, the waveform displayed, and control of the oscilloscope
returned to the user after the timeout has elapsed.

6.

Under Display Settings, touch the Display mode control, and select a
sequence mode display from the pop-up menu.

NOTE

: Once a single acquisition has started, you can interrupt the

acquisition at any time by pressing the

S

INGLE

front panel button a second

time or by pressing the

S

TOP

front panel button. In this case, the segments

already acquired will be retained in memory.

7.

Touch the

S

INGLE

trigger front panel button.