Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 144
LabMaster 10 Zi Oscilloscopes
136
922561-00 Rev A
2.
Touch inside the Title control and provide a value using your
preferred input control method.
3.
If desired, touch inside the Description field, provide information and
touch Save.
The My Notebook Entries list shows the entries you've already stored
on your instrument.
PLEASE NOTE THE FOLLOWING:
When an entry is first created, your new entry is shown along
with the Drawing Toolbar at the top of the screen.
When you select the Done button from the Drawing Toolbar,
your new entry is shown in the My Notebook Entries list.
Flashback Recall
With a stored entry selected on the My Notebook Entries section of the
dialog, touch the Flashback (Recall) button to return your instrument to the
state it was in when the Notebook Entry was saved.
Exit Flashback by touching the Undo Flashback button in the top-right
corner of the screen, or the Auto Trigger front panel control.
PLEASE NOTE THE FOLLOWING: The flashback feature currently recalls the
oscilloscope Setup, and all displayed waveforms. Some of result data is not
recalled, including:
Persistence data - While it is saved in hardcopy and is printed on the
report, it is not recalled during Flashback.
Histogram data - Histograms internally have a 32-bit resolution, but
when stored into a trace file and recalled during flashback they are
clipped to 16-bits.
Floating point waveforms - Certain math operations result in the
creation of floating point waveforms with much higher resolution
than can be stored in a 16-bit waveform file. This extra resolution is
not preserved when traces are recalled using flashback.
Cumulative Measurements - Any measurements on when the Lab
Notebook entry is created are not saved individually in the database