Specifications, Certifications, Emc compliance – Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual
Page 175

Getting Started Manual
922561-00 Rev A
167
Specifications
NOTE
: Specifications are subject to change without notice.
Detailed specifications are maintained on the product Datasheet available
at www.teledynelecroy.com. Always check the website for the most up to
date specifications.
Certifications
This section certifies the instrument’s Electromagnetic Compatibility (EMC),
Safety and Environmental compliances.
EMC Compliance
EC
D
ECLARATION OF
C
ONFORMITY
-
EMC
The oscilloscope meets intent of EC Directive 2004/108/EC for
Electromagnetic Compatibility. Compliance was demonstrated to the
following specifications as listed in the Official Journal of the European
Communities:
EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical
equipment for measurement, control, and laboratory use.
1
Electromagnetic Emissions:
CISPR 11:2003, Radiated and conducted Emissions, Group 1, Class A
2 3
EN 61000-3-2:2006 Harmonic Current Emissions, Class A
EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1
Electromagnetic Immunity
:
EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV
vertical/horizontal coupling planes
4
EN 61000-4-3:2006
RF Radiated Electromagnetic Field, 3 V/m, 80-1000
MHz; 3 V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz
4
EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply
lines, 0.5 kV on I/O signal data and control lines
4
EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE
4