Absolute Volts - Have the vertical offset level indicator move with the
actual voltage level. With this selection, the waveform could move
off the grid.
NOTE
: Regardless of whether you select
Volts
or
Div
, the
Offset
shown in the
Channel Setup dialog always indicates volts. However, when
Div
is selected for the
Offset Control, the offset in volts is scaled proportional to the change in gain,
thereby keeping the division on the grid constant.
Delay Setting constant in - allows you to specify how Horizontal scale
changes are displayed. Choose either:
Time - As the timebase changes, keep the horizontal offset indicator
stationary. With this selection, the trigger point could move off the
grid.
Div(isions) - Have the horizontal offset indicator move with the
trigger point. With this selection, the trigger point remains on the
grid as you increase the timebase.
NOTE
: Regardless of whether you select Time or
Div
, the
Delay
shown in the
Timebase Setup dialog always indicates time.
Reset trigger counter before starting a new acquisition - If available, this
selection clears the trigger counter each time the oscilloscope issues an
ARM acquisition command and applies any trigger Holdoff conditions set on
the Trigger dialog: