beautypg.com

Teledyne LeCroy LabMaster 10Zi Operators Manual User Manual

Page 86

background image

LabMaster 10 Zi Oscilloscopes

78

922561-00 Rev A

As with System Skew calibration, you will see the Fast Edge output
signal appear on the display and delay (skew) measurements will be
made.

7.

Connect the Fast Edge calibration signal to the next input (in this

example, C2), then touch the Next button.

Repeat Step 7 for each channel. As each channel is calibrated, its
trace descriptor appears under the grid display area.

8.

When you have finished all channels, a message indicates all

measurements have been performed. Touch the Finish button. The
ChannelSync™ Calibration dialog shows details about all performed
measurements and the deskew values applied.

If you wish to perform additional ChannelSync calibrations on other
gain settings, repeat the process from Step 5 with a different gain
(V/div) setting and a Reference Correction value entered for that
attenuator.