14 technical data, 1 measuring ranges, resolution, accuracy, 1 ph/orp – YSI TruLab 1320P User Manual
Page 94: Section 14 t, Echnical, Data, Nical
Technical data
TruLab pH/ISE 1320
94
ba76157e01
03/2014
14 Technical data
14.1 Measuring ranges, resolution, accuracy
14.1.1 pH/ORP
Measuring ranges,
resolution
Parameter
Measuring range
Resolution
pH
-2.0 ... +20.0
0.1
-2.00 ... +20.00
0.01
- 2.000 ... + 19.999
0.001
U [mV]
-1200.0 ... +1200.0
0.1
-2500 ... +2500
1
T [°C]
-5.0 ... +105.0
0.1
T [°F]
23.0 ... +221.0
0.1
Manual
temperature input
Parameter
Range
Increment
T
manual
[°C]
-25 ... +130
1
T
manual
[°F]
-13 ... +266
1
Accuracy (± 1 digit)
Parameter
Accuracy
Temperature of the test
sample
pH / range *
-2.0 ... +20.0
± 0.1
+15 °C ... +35 °C
-2.00 ... +20.00
± 0.01
+15 °C ... +35 °C
- 2.000 ... + 19.999
± 0.005
+15 °C ... +35 °C
U [mV] / range
-2500 ... +2500
± 1
+15 °C ... +35 °C
-1200.0 ... +1200.0
± 0.3
+15 °C ... +35 °C
T [°C] / temperature sensor
NTC 30
± 0.1
PT 1000
± 0.1
* when measuring in a range of ± 2 pH around a calibration point
The accuracy values specified here apply exclusively to the meter.
The accuracy of the electrodes and buffer solutions has to be taken
into account additionally.