YSI TruLab 1320P User Manual
Page 64
Ion concentration
TruLab pH/ISE 1320
64
ba76157e01
03/2014
Measuring
CALIBRATION
12.12.2012 12:12:12
#1 0.010 mg/l
20.2 mV 25.0 °C
#2 0.100 mg/l
79.2 mV 25.0 °C
Slp.: 59.0 mV
Sensor +++ (Ion type: Ag)
12.12.2012
12:12
Continue
USB output
5.
Start the measurement with
An entry window appears.
6.
Prepare the sample.
7.
If necessary, add the ISA/TISAB solution to the sample.
8.
Thoroughly rinse the electrode with deionized water.
9.
Immerse the electrode in the sample that was supplemented with blank
value solution.
10.
Using <> <> and , enter the values for the vol-
ume of the sample (Sample volume), the volume of the ISA/TISAB
solution (ISA/TISAB vol.), the volume of the blank value solution (BV
volume) and the concentration of the blank value solution (BV conc.).
11.
Select Continue with <> <> and start the measurement with
.
When the measurement is finished an entry window appears.
i
Immerse sensor in sample
Sample volume
100.0 ml
ISA/TISAB vol.
1.0 ml
BV volume
100.0 ml
BV conc.
1.000 mg/l
Continue
Blank value addition
12.12.2012
12:12
Back
i
Add standard!
Std. volume
100.0 ml
Std. conc.
1.0 mg/l
Continue
Blank value addition
12.12.2012
12:12
Back