YSI TruLab 1320P User Manual
Page 55
![background image](/manuals/310793/55/background.png)
TruLab pH/ISE 1320
Ion concentration
ba76157e01
03/2014
55
11.
Select Continue with <> <> and start the measurement with
.
When the measurement is finished an entry window appears.
12.
Add the standard solution to the sample.
13.
Using <> <> and , enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.) .
14.
Select Continue with <> <> and start the measurement with
.
The measurement result is displayed when the measurement is com-
pleted.
15.
If necessary, start measuring further samples with .
Repeat steps 6 - 14 for all samples.
16.
Terminate the measuring method with
A safety query appears.
17.
Select Yes with <> <>.
18.
Confirm Yes with .
Measurement with the selected method is completed.
i
Add standard!
Std. volume
1.0 ml
Std. conc.
1.0 mg/l
Continue
Standard addition
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Standard addition
mg/l
USB output