beautypg.com

YSI TruLab 1320P User Manual

Page 29

background image

TruLab pH/ISE 1320

pH value

ba76157e01

03/2014

29

Criteria for a stable

measured value

The Stability control function checks whether the measured values are stable
within the monitored time interval.

The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.

1.

Freeze the measured value with .
The [HOLD] status indicator is displayed.
The HOLD function is active.

You can terminate the Stability control function and the HOLD func-
tion with or at any time.

2.

Using , activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the dis-
play of the measured parameter stops flashing.

The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.

You can prematurely terminate the Stability control function manu-
ally with at any time. If the Stability control function is
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.

3.

Using , start a further measurement with stability control.
or
Release the frozen measured value again with or .
The [AR] status display disappears. The display switches back to the
previous indication.

Measured
parameter

Time interval

Stability in the time interval

pH value

15 seconds

: better than 0.01 pH

Temperature

15 seconds

: better than 0.5 °C

This manual is related to the following products: