beautypg.com

8 ion concentration, 1 measuring, 1 measuring the ion concentration – YSI TruLab 1320P User Manual

Page 45: Ion concentration, Measuring, Measuring the ion concentration, 8ion concentration

background image

TruLab pH/ISE 1320

Ion concentration

ba76157e01

03/2014

45

8

Ion concentration

8.1

Measuring

8.1.1 Measuring the ion concentration

NOTE

When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as the values would be incorrect
.

The USB interface is not galvanically isolated.

Stability control

(AutoRead)

& HOLD function

The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display

Incorrect calibration of ion selective electrodes will result in incor-
rect measured values. Calibrate regularly before measuring.

For precise ISE measurements the temperature difference
between measurement and calibration should not be greater that
2 K. Therefore, adjust the temperature of the standard and
measuring solutions accordingly. If the temperature difference is
greater the [TpErr] warning appears in the measured value display.

1.

Connect the ISE combination electrode to the meter.
The pH/ISE measuring window is displayed.

2.

If necessary, select the ISE display (unit, mg/l) with .

3.

If necessary, measure the temperature of the test sample with a ther-
mometer.

4.

Calibrate or check the meter with the electrode.

While no valid calibration is available, e.g. in the delivery condition,
"Error" appears in the measured value display.

5.

Immerse the electrode in the test sample.

ISE

0.157

mg/l

25.0

°C

Ag

12.12.2012

12:12

USB output

This manual is related to the following products: