YSI TruLab 1320P User Manual
Page 52
Ion concentration
TruLab pH/ISE 1320
52
ba76157e01
03/2014
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration.
Calibration record
(example)
Menu item
Setting/func-
tion
Description
Calibration /
Calibration data stor-
age /Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with <><>.
Output the displayed calibra-
tion record to the interface with
Quit the display with
[Back] or .
Switch directly to the mea-
sured value display with
Calibration /
Calibration data stor-
age /
Output to USB
-
Outputs the calibration records to
the interface.
Display
Calibration
record
Magnitude of the slope [mV]
+++
50.0 ... 70.0 or 25.0 ... 35.0
Error
Error
< 50 or > 70
or
< 25 or > 35
Error elimination (see section 13 W
TruLab pH/ISE 1320
Ser. no. 12345678
CALIBRATION ISE
18.01.2013 08:09:10
Sensor
Ser. no. 12345678
Standard 1
0.010 mg/l
Standard 2
0.020 mg/l
Voltage 1
38.5 mV
Voltage 2
58.0 mV
Temperature 1
24.0 øC
Temperature 2
24.0 øC
Ion type
Ag
Slope
64.7 mV
Sensor
+++
_____________________________________