YSI 556 MPS - Quick Start Guide User Manual
Model 556 quick-start guide
Model 556
Quick-Start Guide
This Quick-Start Guide is meant to serve as a quick reference in operating the YSI
Model 556. It is not intended to replace the information found in the Operations
Manual. For your convenience, included in parenthesis for each section, are the section
numbers from the full-length manual where additional information may be found.
INSTALLING THE DO MEMBRANE (3.4)
Note: The DO sensor is shipped with a dry shipping membrane to protect the
electrode. A new membrane cap must be installed before the first use.
1. Prepare the O
2
probe solution according to the instructions on the bottle. After
mixing, allow the solution to sit for 1 hour. This will help prevent air bubbles
from later developing under the membrane.
2. Unscrew and remove the probe sensor guard.
3. Unscrew, remove, and discard the old membrane cap.
4. Thoroughly rinse the sensor tip with distilled or deionized water.
5. Fill a new membrane cap with O2 probe solution. Be very careful not to touch
the membrane surface.
6. Thread the membrane cap onto the sensor, moderately tight. A small amount
of electrolyte should overflow.
7. Screw the probe sensor guard on moderately tight.
MENU FUNCTIONS (2.9 – 2.11)
The Model 556 is set up with a menu-based interface. To navigate through the menus,
use the up and down arrow keys to highlight a desired menu option, then press the
Enter key to open the menu feature. Press the Esc key to return to a previous screen.
The 556 will automatically power on to the Run screen. Press the Esc key to display
the main menu screen.
SETTING THE DATE AND TIME (10.2)
1. Select System Setup from the main menu and then select Date & Time.
2. Highlight Date and press Enter.
3. Use the keypad to enter the correct date and press Enter.
4. Highlight Time and press Enter.
5. Use the keypad to enter the correct military time and press Enter.
6. Press Esc several times to return to the main menu.