YSI TruLab 1320P User Manual
Page 57
![background image](/manuals/310793/57/background.png)
TruLab pH/ISE 1320
Ion concentration
ba76157e01
03/2014
57
10.
Using <> <> and , select the values for the vol-
ume of the sample (Sample volume) and the volume of the ISA/TISAB
solution (ISA/TISAB vol.).
11.
Select Continue with <> <> and start the measurement with
.
When the measurement is finished an entry window appears.
12.
Add the standard (complexing agent or precipitating agent) to the sam-
ple.
13.
Using <> <> and , enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.) .
14.
If necessary, set the ion type for the ion in the standard (Ion type) with
<> <> and .
On selection of the ion type that can be defined (ION):
Using <> <> and , set the valence (Valency) and
molar weight (Molar mass) for the ion in the standard solution.
15.
Select Continue with <> <> and start the measurement with
.
The measurement result is displayed when the measurement is com-
pleted.
16.
If necessary, start measuring further samples with .
Repeat steps 6 - 13 for all samples.
17.
Terminate the measuring method with
A safety query appears.
18.
Select Yes with <> <>.
i
Add standard!
Ion type
S
Std. volume
1.0 ml
Std. conc.
1.0 mg/l
Continue
Standard subtraction
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Standard subtraction
mg/l
USB output