2 safe operation, 1 authorized use, 2 requirements for safe operation – YSI TruLab 1320P User Manual
Page 10: 3 unauthorized use, Safe operation, Authorized use, Requirements for safe operation, Unauthorized use
Safety
TruLab pH/ISE 1320
10
ba76157e01
03/2014
2.2
Safe operation
2.2.1 Authorized use
This meter is authorized exclusively for pH, ORP and ISE concentrations
measurements in the laboratory.
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 14 T
Any other use is considered unauthorized.
2.2.2 Requirements for safe operation
Note the following points for safe operation:
• The meter may only be operated according to the authorized use specified
above.
• The meter may only be supplied with power by the energy sources
mentioned in this operating manual.
• The meter may only be operated under the environmental conditions
mentioned in this operating manual.
• The meter may only be opened if this is explicitly described in this operating
manual (example: Inserting the batteries).
2.2.3 Unauthorized use
The meter must not be put into operation if:
• it is visibly damaged (e.g. after being transported)
• it was stored under adverse conditions for a lengthy period of time (storing
conditions, see section 14 T