TruLab pH/ISE 1320

Ion concentration

ba76157e01

03/2014

49

Continuing with two-

point calibration

7.

Thoroughly rinse the electrode with distilled water.

8.

Immerse the electrode in standard solution 1.

9.

When calibrating without temperature sensor:

Measure the temperature of the standard solution using a thermom-
eter.

Use /[ T

⇅ C ] to select the setting of the temperature.

Use <> <> to set the temperature.

Use /[ T

⇅ C ] to select the setting of the concentration.

10.

Set the concentration of the standard solution with <> <> and
press

.

The standard solution is measured.

The measured value is checked for stability (AutoRead).

11.

Wait for the end of the AutoRead measurement or accept the calibra-
tion value with

.

The calibration display for the next standard solution appears.

ISE

0.0

24.8 °C

12.12.2012

12:12

0.01 mg/

mV

AR

ISE

0.0

24.8 °C

Ag

0.020

mg/l

2

12.12.2012

12:12

Standard

mV

12.

Thoroughly rinse the electrode with distilled water.

13.

Immerse the electrode in standard solution 2.