2 settings for ise measurements, Settings for ise measurements – YSI TruLab 1320P User Manual
Page 73
TruLab pH/ISE 1320
Settings
ba76157e01
03/2014
73
9.2
Settings for ISE measurements
The settings are made in the measuring menu of the ISE measurement. To
open the settings, activate the relevant measuring window in the measured
value display and press the key shortly. After completing the
settings, switch to the measured value display with
The following settings are possible for ISE measurements:
Menu item
Possible setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration data stor-
age /
Display
-
Displays the calibration
record.
Further options:
Scroll through the cali-
bration records with
<><>.
Output the displayed cal-
ibration record to the
interface with
output].
Output all calibration
records to the interface
with
put].
Quit the display with
Switch directly to the
measured value display
with
Calibration /
Calibration data stor-
age /
Output to USB
-
Outputs the calibration
records to the interface.
Calibration/
Serial number (sen-
sor)
-
Entry of the series number
of the connected sensor.
The series number is output
in the calibration record.
Change the contents of
the highlighted position
with <><>.
Go to the next digit with
After theseries number
was completely entered,
confirm with