2 settings for ise measurements, Settings for ise measurements – YSI TruLab 1320P User Manual

Page 73

background image

TruLab pH/ISE 1320

Settings

ba76157e01

03/2014

73

9.2

Settings for ISE measurements

The settings are made in the measuring menu of the ISE measurement. To
open the settings, activate the relevant measuring window in the measured
value display and press the

key shortly. After completing the
settings, switch to the measured value display with .
The following settings are possible for ISE measurements:

Menu item

Possible setting

Description

Calibration /
Calibration record

-

Displays the calibration
record of the last calibration.

Calibration /
Calibration data stor-
age
/
Display

-

Displays the calibration
record.

Further options:

Scroll through the cali-
bration records with
<><>.

Output the displayed cal-
ibration record to the
interface with /[USB
output]
.

Output all calibration
records to the interface
with [USB out-
put]
.

Quit the display with
/[Back] or

ENTER>.

Switch directly to the
measured value display
with .

Calibration /

Calibration data stor-
age
/
Output to USB

-

Outputs the calibration
records to the interface.

Calibration/

Serial number (sen-
sor)

-

Entry of the series number
of the connected sensor.
The series number is output
in the calibration record.

Change the contents of
the highlighted position
with <><>.

Go to the next digit with
/[].

After theseries number
was completely entered,
confirm with .