YSI TruLab 1320P User Manual
Page 33
![background image](/manuals/310793/33/background.png)
TruLab pH/ISE 1320
pH value
ba76157e01
03/2014
33
Continuing with
three- to five-point
calibration
14.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
15.
Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with
The calibration display for the next buffer appears (voltage display).
16.
If necessary, finish the calibration procedure as a two-point calibration
with
The calibration record is displayed.
pH
0.0
24.8 °C
AutoCal TEC
7.000
12.12.2012
12:12
AR
Buffer
mV
17.
Thoroughly rinse the electrode with deionized water.
18.
Immerse the electrode in the next buffer solution.
19.
When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
20.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
pH
180.0
24.8 °C
AutoCal TEC
4.010
12.12.2012
12:12
AR
Buffer
mV