YSI TruLab 1320P User Manual
Page 50
Ion concentration
TruLab pH/ISE 1320
50
ba76157e01
03/2014
Continuing with
three- to seven-point
calibration
Repeat the steps 8 to 13 in the same way with the third and further standard
solutions as necessary. The new calibration values are displayed after the last
calibration step was completed.
14.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use
⇅ C ] to select the setting of the temperature.
Use <> <> to set the temperature.
Use
⇅ C ] to select the setting of the concentration.
15.
Set the concentration of the standard solution with <> <> and
press .
The standard solution is measured.
The measured value is checked for stability (AutoRead).
16.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with .
The calibration display for the next standard solution appears.
17.
Press to continue with three-point calibration.
or
Finish the calibration procedure as a two-point calibration with
The new calibration values are displayed.
ISE
8.9
24.8 °C
12.12.2012
12:12
0.01 m
mV
AR
ISE
20.5
24.8 °C
Ag
0.050
mg/l
3
12.12.2012
12:12
Standard
mV
Based on the calibration data, the calibration curve is determined
in sections, according to the Nernst equation modified by Nikolski.