YSI TruLab 1320P User Manual
Page 62

Ion concentration
TruLab pH/ISE 1320
62
ba76157e01
03/2014
Measuring
7.
Start the measurement with
An entry window appears.
8.
Prepare the standard solution.
9.
If necessary, add the ISA/TISAB solution to the standard solution.
10.
Thoroughly rinse the electrode with deionized water.
11.
Immerse the electrode in the standard solution (complexing agent or
precipitating agent).
12.
Using <> <> and , enter the values for the vol-
ume of the standard solution (Std. volume), the concentration of the
standard solution (Std. conc.) and the volume of the ISA/TISAB solu-
tion (ISA/TISAB vol.).
13.
Select Continue with <> <> and start the measurement with
.
When the measurement is finished an entry window appears.
14.
Add the sample to the standard (complexing agent or precipitating
agent).
15.
Using <> <> and , enter the value for the volume
of the sample (Sample volume).
16.
Select Continue with <> <> and start the measurement with
.
The measurement result is displayed when the measurement is com-
pleted.
i
Immerse sensor in standard
Std. volume
100.0 ml
Std. conc.
1.0 mg/l
ISA/TISAB vol.
1.0 ml
Continue
Sample subtraction
12.12.2012
12:12
Back
i
Add sample!
Sample volume
1.0 ml
Continue
Sample subtraction
12.12.2012
12:12
Back