beautypg.com

4 sample subtraction, Sample subtr – YSI TruLab 1320P User Manual

Page 61

background image

TruLab pH/ISE 1320

Ion concentration

ba76157e01

03/2014

61

8.3.4 Sample subtraction

With the "Sample subtraction" procedure, a known amount of sample is added
to the standard solution (complexing agent or precipitating agent).
The ion concentration in the sample is calculated from the change in potential.
The sample subtraction is one method for the indirect determination of ions. It
is used for the determination of ions for which no direct determination is possi-
ble.

Calibration

1.

Select the measuring method (see section 8.3 S

ELECTING

THE

MEASUR

-

ING

METHOD

, page 53).

2.

If necessary, set the ion type for the ion in the standard (Ion type (sub))
with <> <> and

.

On selection of the ion type that can be defined (ION):

Using <> <> and

, set the valence (Valency) and
molar weight (Molar mass) for the ion in the standard solution.

3.

Select Continue with <> <> and confirm with

.

i

Set the ion type for the standard.

Ion type

S

Continue

Sample subtraction

12.12.2012

12:12

Back

4.

Prepare two standard calibration solutions.

5.

Perform a two-point calibration according to user guidance.

6.

The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.

CALIBRATION
12.12.2012 12:12:12
#1 0.010 mg/l

79.2 mV 25.0 °C

#2 0.100 mg/l

20.2 mV 25.0 °C

Slp.: -59.0 mV
Sensor +++ (Ion type: S)

12.12.2012

12:12

Continue

USB output

This manual is related to the following products: