4 settings for orp measurements, Settings for orp measurements – YSI TruLab 1320P User Manual
Page 72
Settings
TruLab pH/ISE 1320
72
ba76157e01
03/2014
9.1.4 Settings for ORP measurements
The settings for ORP measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the key. After completing the settings, switch to the
measured value display with
In the following table, only those settings are listed that influence the ORP mea-
surement.
Default settings are printed in bold.
Menu item
Possible
setting
Description
Man. temperature
-25 ... +25 ...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Resolution mV
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the
delivery condition (see section 9.4.1
R