6.2.5 Calibration points

Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:

26.

Set the nominal buffer value for the measured temperature with
<><>.

27.

Accept the set calibration value with .

The calibration display for the next buffer appears (voltage display).

28.

Use to finish calibration or
Continue calibrating using the next buffer with .

After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.

Values to be
determined

Displayed calibration data

1-point

Asy

Zero point = Asy

Slope = Nernst slope

(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

Zero point = Asy

Slope = Slp.

3-point to
5-point

Asy
Slp.

Zero point = Asy

Slope = Slp.

The calibration line is calculated by lin-
ear regression.

You can display the slope in the units, mV/pH or % (see section
9.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page 66).

pH

9.958

24.8 °C

ConCal

3

12.12.2012

12:12

Buffer

