YSI TruLab 1320P User Manual
Page 60
Ion concentration
TruLab pH/ISE 1320
60
ba76157e01
03/2014
11.
Select Continue with <> <> and start the measurement with
.
When the measurement is finished an entry window appears.
12.
Add the sample to the standard solution.
13.
Using <> <> and , enter the value for the volume
of the sample (Sample volume).
14.
Select Continue with <> <> and start the measurement with
.
The measurement result is displayed when the measurement is com-
pleted.
15.
If necessary, start measuring further samples with .
Repeat steps 6 - 12 for all samples.
16.
Terminate the measuring method with
A safety query appears.
17.
Select Yes with <> <>.
18.
Confirm Yes with .
Measurement with the selected method is completed.
i
Add sample!
Sample volume
1.0 ml
Continue
Sample addition
12.12.2012
12:12
Back
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Sample addition
mg/l
USB output