YSI TruLab 1320P User Manual
Page 67

TruLab pH/ISE 1320
Settings
ba76157e01
03/2014
67
Calibration/
Serial number (sen-
sor)
-
Entry of the series number of the
connected sensor.
The series number is output in the
calibration record.
Change the contents of the high-
lighted position with <><>.
Go to the next digit with
[].
After theseries number was
completely entered, confirm with
Calibration /Buffer
YSI
ConCal
NIST/DIN
...
Buffer sets to be used for pH cali-
bration (see section 6.2 C
Calibration /One
point calibration
Yes
No
Quick calibration with 1 buffer
Calibration /Cali-
bration interval
1 ... 7 ... 999 d
Calibration interval for the pH elec-
trode (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Calibration /Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst
slope of -59.2 mV/pH (100 x deter-
mined slope/Nernst slope).
Man. temperature
-25 ... +25 ...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahr-
enheit.
All temperature values are dis-
played with the selected unit.
Resolution pH
0.001
0.01
0.1
Resolution of the pH display
Menu item
Possible set-
ting
Description