YSI TruLab 1320P User Manual
Page 42
ORP voltage
TruLab pH/ISE 1320
42
ba76157e01
03/2014
The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.
Manually starting
the stability control
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 9.3.1 S
, page 75) in the
System menu.
Freezing the
measured value
(HOLD function)
With the HOLD function, you can freeze the current measured value. The dis-
played measured value stops changing until you switch the HOLD function off.
Measured parame-
ter
Time interval
Stability in the time interval
ORP voltage
15 seconds
∆
: better than 0.3 mV
Temperature
15 seconds
∆
: better than 0.5 °C
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.
2.
Using
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable measured
value is recognized. The progress bar disappears and the display of the
measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the Stability control function manu-
ally with
prematurely terminated, the current measurement data are output
to the interface without the AutoRead info.
3.
Using
or
Release the frozen measured value again with
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.