2 measuring the temperature, Measuring the temperature – YSI TruLab 1320P User Manual
Page 47
TruLab pH/ISE 1320
Ion concentration
ba76157e01
03/2014
47
8.1.2 Measuring the temperature
For reproducible ion-selective measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
Temperatur
e sensor
Resolution of the
temp. display
Temp. measurement
yes
0.1 °C
Automatic with temperature sensor
-
1 °C
Manual
1.
Measure the current temperature of the test sample.
2.
Set the temperature value with <><>.
or
In the /ISE/Man. temperature menu, set the tempera-
ture value with <><>.