4 calibration standards, 5 calibration data, Calibration standards – YSI TruLab 1320P User Manual
Page 51: Calibration data
TruLab pH/ISE 1320
Ion concentration
ba76157e01
03/2014
51
8.2.4 Calibration standards
Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.
8.2.5 Calibration data
Displaying the
calibration data
The calibration record of the last calibration is to be found under the menu item,
/ Calibration / / Calibration record. To open it, press the
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with
Displaying the
calibration data
storage
The calibration records of the last calibrations (up to 10) are given in the menu,
/ ISE / Calibration / Calibration data storage.
Select the unit of the standard solution and measurement result in
the ISE setup/Unit menu.
Standard solution (Std 1 - 7) Values
Unit [mg/l]
0.010 ... 500,000
Unit [mol/l]
0.100 ... 5,000 µmol/l
10.00 ... 5,000 mmol/l
Unit [mg/kg]
0.010 ... 500,000
Unit [ppm]
0.010 ... 500,000
Unit [%]
0.001 ... 50,000
The measurement precision is also dependent on the selected
standard solutions. Therefore, the selected standard solutions
should cover the value range expected of the subsequent concen-
tration measurement.
If the measured electrode potential is outside the calibrated range,
the [ISEErr] warning is displayed.