Ph 1 – YSI TruLab 1320P User Manual
Page 32
pH value
TruLab pH/ISE 1320
32
ba76157e01
03/2014
Continuing with two-
point calibration
5.
Thoroughly rinse the electrode with deionized water.
6.
Immerse the electrode in the first buffer solution.
7.
When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
8.
Start the measurement with
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.
Wait for the end of the measurement with stability control or accept the
calibration value with
The calibration display for the next buffer appears (voltage display).
10.
If necessary, finish the calibration procedure as a single-point calibration
with
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.
pH
1
-180.0
24.8 °C
AutoCal TEC
12.12.2012
12:12
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
12.12.2012
12:12
AR
Buffer
mV
11.
Thoroughly rinse the electrode with deionized water.
12.
Immerse the electrode in the second buffer solution.
13.
When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.