beautypg.com

Ph 1 – YSI TruLab 1320P User Manual

Page 32

background image

pH value

TruLab pH/ISE 1320

32

ba76157e01

03/2014

Continuing with two-

point calibration

5.

Thoroughly rinse the electrode with deionized water.

6.

Immerse the electrode in the first buffer solution.

7.

When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.

8.

Start the measurement with .
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

9.

Wait for the end of the measurement with stability control or accept the
calibration value with .

The calibration display for the next buffer appears (voltage display).

10.

If necessary, finish the calibration procedure as a single-point calibration
with .
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.

pH

1

-180.0

24.8 °C

AutoCal TEC

12.12.2012

12:12

Buffer

mV

pH

-180.0

24.8 °C

AutoCal TEC

10.011

12.12.2012

12:12

AR

Buffer

mV

11.

Thoroughly rinse the electrode with deionized water.

12.

Immerse the electrode in the second buffer solution.

13.

When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.

This manual is related to the following products: