YSI TruLab 1320P User Manual
Page 90
What to do if...
TruLab pH/ISE 1320
90
ba76157e01
03/2014
No stable measured
value
Obviously incorrect
measured values
Cause
Remedy
pH electrode:
– Junction contaminated
– Clean the junction
– Membrane contaminated
– Clean the membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Temper if necessary
Electrode + test sample:
– Conductivity too low
– Use a suitable electrode
– Temperature too high
– Use a suitable electrode
– Organic liquids
– Use a suitable electrode
Cause
Remedy
pH electrode:
– pH electrode unsuitable
– Use a suitable electrode
– Temperature difference between
buffer and test sample too great
– Adjust the temperature of buf-
fer or sample solutions
– Measurement procedure not suit-
able
– Follow a special procedure