YSI TruLab 1320P User Manual
Page 65
TruLab pH/ISE 1320
Ion concentration
ba76157e01
03/2014
65
12.
Add the standard solution to the sample.
13.
Using <> <> and , enter the values for the vol-
ume of the standard solution (Std. volume) and concentration of the
standard solution (Std. conc.).
14.
Select Continue with <> <> and start the measurement with
. The measurement result is displayed when the
measurement is completed.
15.
If necessary, start measuring further samples with .
Repeat steps 6 - 14 for all samples.
16.
Terminate the measuring method with
A safety query appears.
17.
Select Yes with <> <>.
18.
Confirm Yes with .
Measurement with the selected method is completed.
ISE
0.321
24.8 °C
Ag
12.12.2012
12:12
Blank value addition
mg/l
USB output