2 settings for ise measurements, Settings for ise measurements – YSI TruLab 1320P User Manual
Page 73

TruLab pH/ISE 1320
Settings
ba76157e01
03/2014
73
9.2
Settings for ISE measurements
The settings are made in the measuring menu of the ISE measurement. To 
open the settings, activate the relevant measuring window in the measured 
value display and press the  key shortly. After completing the 
settings, switch to the measured value display with 
The following settings are possible for ISE measurements:
Menu item
Possible setting
Description
Calibration / 
Calibration record
-
Displays the calibration 
record of the last calibration.
Calibration / 
Calibration data stor-
age /
Display
-
Displays the calibration 
record.
Further options:
Scroll through the cali-
bration records with 
<><>.
Output the displayed cal-
ibration record to the 
interface with 
output].
Output all calibration 
records to the interface 
with 
put].
Quit the display with 
Switch directly to the 
measured value display 
with 
Calibration /
Calibration data stor-
age /
Output to USB
-
Outputs the calibration 
records to the interface.
Calibration/
Serial number (sen-
sor)
-
Entry of the series number 
of the connected sensor. 
The series number is output 
in the calibration record.
Change the contents of 
the highlighted position 
with <><>.
Go to the next digit with 
After theseries number 
was completely entered, 
confirm with 