Gas sampling and gas conditioning technology

2-1.1.4-ME

W A R N I N G !


Aggressive condensate possible.

Wear protective glasses and protective clothes!

W A R N I N G !


For works during operation:

High surface temperatures!

Any contact may lead to burnings.

Wear protective gloves and protect the probe against unauthor-
ized access!

16.1 REPLACEMENT OF THE STANDARD FILTER ELEMENT AND CONTROL OF THE

SEALS

Shut the ball valve (if present). Purge the probe if toxic gases have been used!

Remove the protective cover.

Take out the filter receiving part as described in chapter 12.1.

Unscrew the filter knurled screw I and replace the filter element J.

Control the filter element seals K and replace them if necessary.

Control the flat seal of graphite inside the lid and replace it if necessary.

Clean the filter space.

Push eventually through the sample tube to remove any deposits.

Caution! Risk of fracture when using sample tubes of aluminium oxide.

Then insert the filter receiving part again. Set the handle C in position E and tighten the lid

again by using the locking handle A.

Put on the protective cover.

Open the ball valve.