M&C TechGroup CSS-V2 Data sheet User Manual
Gas conditioning unit series css, Special features, Application
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
14.1
Embracing Challenge
Compact high-performance construc-
tion with compressor cooler
Individually configurable
Completely pre-installed
One or two gas streams
19”- or wall mounting housing
Low maintenance and self-monitoring
Outlet dew point adjustable from +2 °C
to +7 °C
Dew point stability ± 0,1°C
Operational in 10 minutes
Light weight
Optimum reliability
Special Features
Gas Conditioning Unit series CSS®
Version CSS-V1 for max. 1 x 250Nl/h gas flow rate,
Version CSS-V2 for max. 2 x 150Nl/h gas flow rate,
for 19”- or wall mounting
CSS-V2
Application
The units CSS-V.. provide a complete pre-
installed compact high-performance sample
gas conditioning for continuous use. Due to a
big variety of additional options the gas con-
ditioning units excellently can be adapted to
the most diverse requirements of gas analysis
technique.
Its compact construction only takes up little
space. The CSS-V units are ready for use in a
few minutes. This makes time-consumingpro-
curement of individual components and
assembly superfluous.
Description
All components of the gas conditioning unit
are mounted in a compact sheet steel case.
Filter, flowmeter and peristaltic pump are
placed into the front panel and thus assure a
very easy maintenance.
The compressor gas cooler is equipped with
one or two heat exchangers of Duran glass,
stainless steel SS316Ti or PVDF.
The fine filter FPF-2-0,3GF (0,3µm filter fine-
ness) installed in front of the sample gas
pump N3/5/9 KPE ensures the necessary
removal of solid particles. The condition of
the filter element is visible from outside.
The downstream installed sample gas pump
N3/5/9 KPE is available with 3 pump capaci-
ties.
The flow meter FM40 with corresponding
measuring range installed in the sample gas
outlet can be equipped with a flow alarm
sensor FA-20mo and the adequate electronic.
The unit contains a temperature alarm con-
tact that is activated in case the temperature
differs more than +/-3 °C from the set value (+
5°C) adjusted at the factory and switches off
the sample gas pump if existing.
The condensate occurred is removed con-
tinuously via the peristaltic pump type SR25.2.
The ventilation grids in the sidewalls ensure
that the equipment is sufficiently ventilated.
A liquid alarm sensor is integrated in the filter
FPF-2-0,3GF to protect the downstream ana-
lysers against liquid irruptions and to increase
the operating safety of the whole system. Via
one or two potential free contacts for the
summery alarm (cooler temperature, gas flow,
liquid) an alarm is indicated and the sample
gas pump, if existing, is switched off.
Technical specifications and illustrations are without
obligation, subject to modifications. 11.08/11.12