beautypg.com

M&C TechGroup PMA 10S Data sheet User Manual

Oxygen analyser series pma, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

13.2

Embracing Challenge

ƒ

ƒ

Discontinuous oxygen purity measure-
ment within the expanded range of 99
to 100 volume % O2

ƒ

ƒ

Portable and rugged instrument, light-
weight

ƒ

ƒ

Analogue and digital linear
indication

ƒ

ƒ

Physical measuring principle

ƒ

ƒ

Small dead volume, fast response time

ƒ

ƒ

High accuracy and reliability

Special Features

Oxygen Analyser Series PMA®

Portable version PMA 10S
for oxygen purity measurement

PMA 10S

Application

Due to the extremely fast response time of

the patented M&C magneto-dynamic meas-

uring cell with no stagnant volume as well

as the negligible cross sensitivity from other

sample gas components, the portable M&C

oxygen analyser PMA 10S does not normally

limit its use for a wide variety of applica-

tions where oxygen purity measurements are

essential.

Description

The PMA 10S oxygen purity analyser, being an

extended version of the PMA 10, is provided

with an electronic suppression system which

enables oxygen purity measureents within

the expanded 99-100 vol.% range.
This measuring range is displayed on the

analogue meter whilst the digital indicator

remains for the 0-100% reading.
The electronic zero suppression system may

be selected by means of a push button switch

which is located on the control panel of the

instrument.
The PMA 10S analyser can normally be used

also for its standard PMA10 functions. Oxygen

purity measurements could only be made

discontinuously. The measuring accuracy of

the analyser can be maintained when calibra-

tion of the instrument with a certified gas

takes place before the measurement proce-

dure in order to avoid influence of ambient

temperature and pressure. A measurement of

1 vol.% can be easily realized whilst the maxi-

mum value remains at 100 vol.%. Other meas-

urement within this range can be performed.
– Further technical data see leaflet PMA10,

13.1. –

Measuring principle of M&C ox-

ygen analyser

The PMA 10S utilises the paramagnetic

principle of operation to measure oxygen

concentrations. The analyser measures the

paramagnetic susceptibility of the oxygen

in the sample gas by means of the M&C

magneto-dynamic measuring cell. The physi-

cal propertys which distinguishes oxygen

from other gases is its paramagnetism. It is

significantly higher comparing to other com-

mon gases. This operation principle is one of

the most accurate and reliable procedures

to determine the oxygen concentration in a

gas mixture from 0 to 100 Vol.%. The robust

M&C cross-flow cell has no stagnant volume.

Advantages are the fast response time, the

flow rate up to 60 l/hr, the small volume of

2 ml, the extremely low drift, the absolute

linearity and the negligible cross sensitivity

against other sample gas components. With

a proper sample conditioning and pressure,

the M&C cell will never need replacing. The

dumbbell with a small mirror at its centre is

mounted in a strong inhomogenous magnet-

ic field. The paramagnetic oxygen strength-

ens the forces on the diamagnetic dumbbell

and causes a shifting which is detected by

a system consisting of light beam, mirror

and a photo cell. A compensation current is

induced via the feedback coil on the dumb-

bell and leads to a reset of the dumbbell into

its zero-position.
The required current is linearly proportional to

the oxygen concentration.

Technical specifications and illustrations are without