beautypg.com

M&C TechGroup FPF-0,1 Operator's manual User Manual

Instruction manual, Fpf-0,1, Ultra-fine filter series

background image

Subject for technical modifications Reproduction

of

this

document or its content is not allowed without permission of M&C

M&C TechGroup Germany GmbH

z Rehhecke 79 z 40885 Ratingen z Germany z Tel.: (0)2102/935-0 z Fax: (0)2102/935-111 z www.mc-techgroup.com

© M&C 06.97/07.02

Gas sampling and gas conditioning technology

5 - 0.1 ME

Instruction Manual

Ultra-fine filter series

FP

®

,

v

ersion

FPF-0,1

for panel-mounting

Introduction

The M&C panel-mounting filter FPF-0,1 is equipped with an extra fine glass-fibre filter element with a filter
fineness of 0,1 µm. This guarantees an optimised protection of the analysers against solids.

Description

The filter FPF-0,1 consists of the following components:
c filter glass, d filter element holder, e filter element, f O-ring, g filter body, h hexagon screws and i front-
ring.
At the back of the FPF-0,1 filter the G1/8“i

(DIN ISO 228/1)

connections for gas in- and outlet are located. In- and

outlet are marked accordingly. The filter element

e is fixed with the filter element holder d on the filter body g.

The filter glass

c is screwed in the filter body g and covers the filter area. The filter glass presses on the O-ring

f which guarantees the tightness of the device. The front-ring i with it’s hexagon screws fixes the filter FPF-0,1
in the front panel.

Function

The sample gas enters the filter via the gas inlet from the outer side of the filter element to the inner side and
leaves the filter via the outlet connection.

Maintenance
Before the maintenance work is carried out, it is necessary that the specific safety procedures
pertaining to the system and operational process be observed!

Maintenance intervals depend on the

process conditions and have to be determined specifically. The maintenance work is focused on the check of the
filter element respectively the change of the element.

Change of the filter element

Please carry out the following steps:
• before changing the element shut off the gas way to the filter and if necessary sweep the filter with inert gas;
• unscrew by hand the filter glass anticlockwise (the direction is marked on the front-ring);
• unscrew by hand the filter element holder anticlockwise;
• dismantle the filter element;

Check the right position of the o-ring before assembling the filter

• clean the filter housing respectively the filter glass if necessary and re-install the new filter element in the

opposite

turn

Spare parts

Wear, tear and replacement part requirements depend on specific operating conditions.
The recommended quantities are based on experience and are not binding.

Ultra-fine filter version FPF-0,1 (C) consumable parts, (R) recommended spare parts and (T) spare parts


Recommended quantity

for operation [years]

Article-No.

Description

V/E/T

1

2

3

90 F 0009

Glass-fibre filter element F-0,1GF 50, fineness 0,1µm,

e

C

12 24 36

90 F 0097

Glass-fibre filter element F-2GF 50, fineness 2µm,

e

C

12 24 36

90 F 0096

Stainless steel filter element F-3SS50, 3µm*,

e

C

2 4 6

90 F 0118

Filter glass F-45,

c

R

1 2 2

90 F 0043

O-ring of Viton

®

, 41 for FPF-0,1 until 5.93,

f

R

1 2 2

90 F 0044

O-ring of Viton

®

, 35 for FPF-0,1 from 6.93,

f

R

1 2 2

90 F 0095

PVDF filter element holder FPF-GF,

for elements 0,1GF50 / 3SS 50

d

S

- 1 1

90 F 0094

PVDF filter element holder FPF-P,

for elements 2GF 50 *

d

S

-

1 1

*Option

Installation, maintenance, monitoring and any repairs may only be done by authorised personnel with respect to
the relevant stipulations.