M&C TechGroup SP2000H320_DIL... Data sheet User Manual
Gas sample probe series sp, Special features, Application

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
5.1
Embracing Challenge
Based on probe SP2000H/....
Electrically heated up to 180 °C or 320 °C
External critical orifice is heated
Including a dilution gas pre heater
No dew point problems
Dilution ratios from 10:1 up to 500:1
With check gas connection
Independent of the process temperature
Easy operation and maintenance
Special Features
Gas Sample Probe Series SP®
Version SP2000H/DIL... and SP2000H320/DIL...
Gas Sample Dilution Probe
SP2000-H/DIL
Application
The non-heated or electrically heated M&C
dilution unit DIL-1/H is used in analysis tech-
nique to dilute the sample gas or compo-
nents in the sample for example in case of
toxic gases, moisture measurement or emis-
sion measurement.
The M&C dilution unit is based on the func-
tional dilution technique which is proved
since years in the M&C gas sample probe
SP2000-H/DIL.
Description
The M&C dilution units DIL-... are mounted on
a plate for wall mounting. The version DIL-1 is
not heated and works at ambient tempera-
ture. The version DIL-1/H is electrically heated
up to 180°C . It is additionally equipped with
a lagged cover (version for 320°C on request).
A capillary thermostat controlls the
temperature, adjustable from 0-180°C.
A high temperature limiter and a low tem-
perature alarm is included.
The connection of the heated lines is
made without cold bridges in the heated
part. Before the dilution gas enters the
dilution unit, it is heated up to operat-
ing temperature via a gas pre-heater. In
order to protect the dilution part against
contaminations, internal protective fil-
ters are installed for the sample gas and
the dilution gas stream. Calibration gas can be
feeded via the integrated connection.
A precision pressure controller with pressure
gauge is used for adjustment of the necessary
admission pressure of the dilution gas. Via a
vacuum pressure gauge, the function of the
dilution injector is controlled.
Both, pressure controller and pressure gauges
have to be ordered separately. Two versions
are available: Set -A (-A1) is mounted directly
on the mounting plate and control panel type
-S (-S1) is suitable for external 19» rack mount-
ing. A shut off valve and a flowmeter for
adjustment of the calibration gas are included
in the version -S (-S1).
The dilution unit can realize dilution ratios
of 10:1 to 500:1. In case of high dilution
ratios, a respective small quantity of gas
is sampled from the process. Optionally
a bypass injector -B (Option I) is available
integrated right in front of the dilution unit
in order to shorten the response time in case
of operation with atmospheric pressure or
low pressure. In case of option I, an additional
pressure controller is included in both mount-
ing versions.
The construction of the dilution unit guaran-
tees easy maintenance and operation inde-
pendent from process temperatures.
Technical specifications and illustrations are without
obligation, subject to modifications. 08.99/06.06