Application

Typically, the FP-BF humidifier is used when

the humidity of the sample gas and the

calibration gas differs and the analyser shows

a significant cross-sensitivity against water

vapour. Before entering the cooler, the zero

gas passes the humidifier and leaves the

cooler saturated at a stable dew point as well

as the sample gas. Test gases with water-

soluble components must not be passed via

the humidifier.

Description

The M&C FP-BF humidifier consists of our

standard filter components. With the F-120G

or F-120G-D glass body, the FP-BF humidifier

has only a volume of 70 ml.
The gas passes a bubbler that intensifies

the humidifying effect. The glass humidifier

body enables the control of the liquid level

without dismantling the body. No tools are

required for refilling the humidifier. The opti-

mal position of the o-ring guarantees tight-

ness between head and glass body.
The humidifier can be rotated about 180° in

the holder for wall mounting. This allows a

flexible adaptation of in- and outlet to local

conditions.
For special applications, the FP-BF-GL25 can

be provided with a GL25-12 connection at the

bottom of the glass body.
Overpressure at the outlet or underpressure

at the inlet of the humidifier must be avoided.

Otherwise the liquid will be displaced to the

gas inlet.