beautypg.com

M&C TechGroup DIL-U Data sheet User Manual

Sample dilution system version dil-u, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

5.3

Embracing Challenge

ƒ

ƒ

Lowest dilution ratio 1:1

ƒ

ƒ

Ambient air as dilution gas possible

ƒ

ƒ

Low dilution gas consumption

ƒ

ƒ

No influence of barometric pressure

1

ƒ

ƒ

Compact design

ƒ

ƒ

Gas sample probe SP2000H/...

ƒ

ƒ

Incorporated pre heater for dilution gas

ƒ

ƒ

Test gas connection at the probe inlet

ƒ

ƒ

Completely heated up to 180°C

1

varying process pressure must be taken into
consideration

Special Features

Sample dilution system version DIL-U

consisting of dilution probe SP2000H/DIL-U/P
and dilution pump unit DIL-U/A (N)

DIL-U/A(N)

Application

The electrically heated M&C dilution system

DIL-U is applicable for such processes, where

the measuring procedure or the handling of

the process gas requires the dilution of the

sample respectively the component to be

measured in a range of 100:1 up to 1:1.
Typical applications:

toxic gases above TLV value

explosive gas mixture above LEL

adapting the concentration of the sam-

ple components to the measuring range

of the analyser

measuring possibility at low sample flow

rates from 1Nl/hr

moisture measurement in flue gas with

IR-analyser

The M&C dilution probe (part of this system)

is based on the well tested modular designed

M&C gas sample probe SP2000H. This enables

the user to adapt the probe to nearly every

application, using for example special filter

techniques or special materials.

Description

The M&C dilution system DIL-U consists of the

dilution probe SP2000H/DIL-U/P and the dilu-

tion pump unit DIL-U (A/N).
The dilution unit, including the critical ori-

fices, is integrated in the sample probe outlet,

heated up with the probe to a stable temper-

ature. The incorporated pre heater heats

up the dilution gas to the probe temperature.

Both steps avoid that the temperature of

sample gas decreases under the dew point.

The test gases for the system calibration enter

the sample probe via an integrated valve.
A vacuum pump integrated in the pump unit

DIL-U sucks the required vacuum, which is

necessary for the function of the critical orific-

es. A pressure gauge indicates the function of

the dilution system. The flowmeters indicate

the analyser and bypass flow 1. A flow alarm

is available as well.

For test procedure with zero and span gas a

3 way ball valve and a flowmeter is installed.
Version DIL-U/A is suitable for ambient air as

a dilution gas and version DIL-U/N for bottle

gas. Bottle gas requires a pressure regulator,

a pressure gauge and a flowmeter which can

be installed optional in the pump unit.
If the dew point of the diluted sample is

higher then ambient temperature an option-

al integrated M&C cooler, ECP1000G/SR25,

reduces it to +5 °C.
High dilution ratios cause a very low sam-

ple flow. Therefore it is necessary to use an

additional sample bypass 2 to reduce the

response time. This will be done with the

option DIL-U/B with an additional pump and

flowmeter and the cooler as an option.
The heated M&C sample line(s) 3/4-N/M/H

between the probe and the pump unit can

be controlled via an temperature controller

703, built into the pump unit.
The design of the dilution system DIL-U guar-

antees easy maintenance.

Technical specifications and illustrations are without

obligation, subject to modifications. 12.99/06.06