M&C TechGroup SP2600-H_C_I_BB_F_0,1GF190 Data sheet User Manual
Gas sample probe series sp, Special features, Application

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.5
Embracing Challenge
Filter element with particularly large
filter surface
Optimum operational reliability
Universal applicability
Adaption to nearly all process condi-
tions due to its compact and modular
design
Easy mounting and maintanance
Small volume, fast response time
Integrated possibility for back-flushing
and shut off the sample outlet
Easy operation and maintenance
Special Features
Gas Sample Probe Series SP®
Version SP2600-H/C/I/BB/F/1K190 and
SP2600-H/C/I/BB/F/0,1GF190
SP2600-H/C/I/BB/F/1K190
Application
The probes M&C type SP2600-H/C/I/BB/F...,
based on type SP2000-H (see data sheet
2.2.2), are used for continuous gas sampling
in processes with gases of high dust content,
high temperatures and/or high humidity.
Due to the particularly large filter surface and
the possibility to back purge the filter ele-
ment, this type of probes achieve a very long
service life especially in those cases where
no pre-filters can be used. For this purpose, a
back purge valve and a pneumatic isolation
valve for the measuring gas outlet are inte-
grated inside the probe.
Version SP2600-H/C/I/BB/F/0,1GF is suitable
for dusts with extremely fine particles of up
to 0,1 µm grain size. Version SP2600-H/C/I/
BB/F/1K200 can be back purged very effi-
ciently due to its double-ply constructed filter
element with an external filter membrane of
1 µm filter porosity.
Description
The sample probes are designed for easy
installation, reliable operation, trouble free
maintenance and a versatile application.
Depending on the application, different
sample tubes (see data sheet 2-1.1.0.6), not
included in the scope of delivery of the probe,
are screwed into the thread (G3/4”i) of the
filter housing.
The large surface filter element of ceramic
or glass fiber is placed in a housing with low
stagnant space outside the process.
The probes are constructed in such a way
that changing the filter element does not
require the sample line to be dismounted
and, therefore, the contamination of the clean
gas is avoided.
The special design of the heating element of
the SP2600... (with protective hood) permits
controlled heating of the complete filter
housing including the mounting flange up to
180 °C (Version /H320 up to 320 °C).
This ensures reliable operation external to the
process preventing the temperature falling
below the dew-point.
The temperature of the standard probe is
controlled by an integrated compact design
capillary sensor thermostat with an excess
temperature limit switch and an alarm func-
tion for temperature failure in compact
arrangement directly to the probe.
Feeding the test gas is possible via an inte-
grated check valve.
Additional functions of the probe SP2600-
H/C/I/BB/F...:
•
Calibration gas will be injected into the
probe through a check valve /C directly
to the sample outlet. No calibration gas
is lost into the stack.
•
An isolation valve /I shuts off the sam-
ple outlet from the internal filter area.
•
Through a high flow rate check valve /
BB/F, which is fixed to the filter housing
wall, blow back of the incorporated
ceramic filter element will be done incl.
the filter area and the insitu probe tube
or the pre-filter.
Technical specifications and illustrations are without
obligation, subject to modifications. 03.03/02.08