Anritsu Series MS278XB User Manual
Page 252

Operation Verification
Measurements
5-96
MS278XB OM
10. Record the maximum Spurious amplitudes in the appropriate column of Table 5-14 based on the
frequency offset from the carrier frequency.
Note: Any spurs found should be verified not to originate from the signal source.
Table 5-14. Input-related Spurious Responses Test
Center
Frequency
(MHz)
Source
Frequency
(MHz)
Spur Amplitude
f
<
300 kHz from
Carrier
Specification
(dBc)
Spur Amplitude
f
≥300 kHz from
Carrier
Specification
(dBc)
10
10
–73
–80
12.5
12.5
–73
–80
16
16
–73
–80
21.5
21.5
–73
–80
32
32
–73
–80
142.5
142.5
–73
–80
500
500
–73
–80
650
650
–73
–80
820
820
–73
–80
950
950
–73
–80
999
999
–73
–80
1020
1020
–73
–80
2640
2640
–73
–80
3000
3000
–73
–80
3500
3500
–73
–80
4000
4000
–73
–80
4500
4500
–73
–80
5000
5000
–73
–80
5500
5500
–73
–80
6000
6000
–73
–80