Displaying trace modes, Displaying trace modes -26 – Anritsu Series MS278XB User Manual
Page 182
Measurement Examples
Measurements
5-26
MS278XB OM
Displaying Trace Modes
The following example shows the display of a 1 GHz carrier using the MaxHold, Average, and MinHold trace
states:
1. Preset the analyzer using the Preset button.
2. Set the center frequency to 1 GHz and the span to 300 kHz.
3. On the Trace Config menu, set Trace 1 to the MaxHold state.
4. Select Trace 2 as the active trace and set it to the Average state with 5 averages.
5. Select Trace 3 as the active trace and set it to the MinHold state.
The display should look like the one in the figure below:
Figure 5-7.
Displaying Trace Modes