Operation Verification

Measurements

5-90

MS278XB OM

10. Subtract the new reading from the marker reading at –70 dBm Reference Level setting.
11. Verify that the deviation is within specifications.
12. Repeat steps 8 to 12 for other Reference Level settings listed in Table 5-11.

Table 5-11. Reference Level Switching Uncertainty Test

MS2781B

Reference Level Setting

MS2781B

Marker Reading

Deviation from

–70 dBm Reference Level

(dB)

Specifications

–70 dBm

0

N/A

–60 dBm

0.1 dB

–50 dBm

0.1 dB

–40 dBm

0.1 dB

–30 dBm

0.1 dB

–20 dBm

0.1 dB

–10 dBm

0.1 dB

0 dBm

0.1 dB