Measurement Computing WaveBook rev.5.3 User Manual
Page 64

Example 2: WaveBook/516E System with a WBK17 and a WBK40
Device Interface Side Channel Input Side
WaveBook/516E System
with a WBK17 and a WBK40
WBK40
WBK17
WaveBook/516E
Power
Ethernet
Power
Lines and Cables Used in Example, excluding channel input connections.
1. Power Supply - To the WaveBook/516E POWER IN DIN 5 connector, +10 VDC to +30 VDC,
typically from a TR-40U power adapter.
2. Power Supply - To the WBK17’s POWER IN DIN 5 connector, +10 VDC to +30 VDC, typically
from a TR-40U power adapter.
3. Power Supply – From the WBK17’s POWER OUT DIN 5 connector, +10 VDC to +30 VDC, to the
WBK40’s POWER IN DIN 5 connector via a CA-115 Power Cable.
4. Ethernet – From the WaveBook/516E ETHERNET jack to a 10/100BaseT Ethernet Network or to
the ETHERNET jack on a PC. Applicable Ethernet patch cables (1.5-foot or 7-foot) are CA-242 and
CA-242-7, respectively.
5. Synchronization – From a WaveBook/516E SYNC jack to a WBK40 SYNC jack. The connection
is made via a CA-74-1 cable.
6. Expansion – From WaveBook/516E EXPANSION PORT 1 to the WBK40’s PARALLEL PORT
connector. The connection is made via a 25-pin CA-35-12 cable.
7. Expansion Control – From WaveBook/516E EXPANSION CONTROL to the WBK17’s
EXPANSION CONTROL IN. Connection is via a 15-pin CA-129 cable.
8. Expansion Signal – From the WBK17’s EXPANSION SIGNAL OUT to the WaveBook/516E’s
EXPANSION SIGNAL IN. Connection is via a BNC coaxial CA-150 cable.
9. Digital Expansion – This connection is made on the channel input side of the units. The cable runs
from the WaveBook/516E’s DIGITAL I/O / EXTERNAL CLOCK / TTL TRIGGER connector to
the WBK17’s IN-DIGITAL EXPANSION connector (HD26 type). The connection is made via a
CA-217 Digital Expansion Cable.
3-4 System Examples
897895
WaveBook User’s Manual