7-8 Troubleshooting

979194

WaveBook User’s Manual -

IO version

8. WBK15 channel readings are wrong.

Possible Causes

Details and Possible Solutions

> Note: The preceding table also applies to WBK15.

The wrong 5B

type has been

selected.

Check that the correct 5B Module type is selected in WaveView. Selection is made by clicking on the

applicable WBK15 channel’s Range / Mode cell in WaveView’s Channel Configuration screen. This

brings up a pull down list from which you can select the proper 5B module, for example, a 5B30-06 or

a

5B47-B-11
The WBK15 section of the WBK Options Manual includes software setup information pertaining to 5B

modules. A PDF version of the document is included on the data acquisition CD.

The wrong

scaling

is in effect.

WaveView automatically selects the scale related values m and b (in the y = mx + b equation) according

to the type of 5B module; however, a user can easily change these values to customize the engineering

units. Refer to the technical information for the specific 5B Module to ensure that the proper

correlation exists between the input sensor’s units and 5B module’s output scale. In regard to

customizing the scale via WaveView, refer to the WBK15 Software Setup section in the WBK Options

Manual.

There is

insufficient

power to the 5B

module.

5B Modules require +5VDC power and some minimum amount of available current, which is typically

supplied by the WBK15. In cases where strain gage modules are used, ensure that no overloading of

the excitation output is occurring through bridge circuits.

The

thermocouple

leads are

reversed.

If readings appear to be negative when they should be positive, verify that the thermocouple leads are

not reversed.

The load cell

excitation source

is overloaded or

not connected

properly.

Some 5B Module types, for instance strain gage conditioners, provide an excitation source for the

connected sensor. Check the specifications for the module and make sure that the excitation source is

properly connected and not over-loaded.

9. WaveView Scope stops.

Possible Causes

Details and Possible Solutions

Auto-Rearm is

disabled.

If the WaveView Scope acquires and displays data but then stops running, it is most likely due to the

Auto-rearm feature being disabled. If you desire to view data continuously click the auto-rearm button

in the scope window, or select auto-rearm in the Scope window’s Acquire pull-down menu.
Refer to the section, WaveView Scope Window, in the WaveView & WaveCal document for additional

information.